    Publication Date: 2014-02-03
    Description: The validation of a fully automated dissolved Ni monitor for in situ estuarine studies is presented, based on adsorptive cathodic stripping voltammetry (AdCSV). Dissolved Ni concentrations were determined following on-line filtration and UV digestion, and addition of an AdCSV ligand (dimethyl glyoxime) and pH buffer (N-2-hydroxyethylpiperazine-N'-2-ethanesulphonic acid). The technique is capable of up to six fully quantified Ni measurements per hour. The automated in situ methodology was applied successfully during two surveys on the Tamar estuary (south west Britain). The strongly varying sample matrix encountered in the estuarine system did not present analytical interferences, and each sample was quantified using internal standard additions. Up to 37 Ni measurements were performed during each survey, which involved 13h of continuous sampling and analysis. The high resolution data from the winter and summer tidal cycle studies allowed a thorough interpretation of the biogeochemical processes in the studied estuarine system. Copyright (C) 1998 Elsevier Science B.V.
    Type: Article , PeerReviewed