Summary A square-wave polarograph [3] has been used to determine trace amounts of copper in potable water and in Iowa River water. By using a combination of sodium chloride (1.0 M) and perchloric acid (0.07 M) as a supporting electrolyte under acidic conditions (pH〈1), 2 ppb copper could be determined in the samples without and with preconcentration. Copper in the samples was determined with reproducibility of ± 0.13 ppb and so this technique is more reliable than a more used determination of complex salts by atomic absorptionspectrophotometry.
Notes:
Zusammenfassung Ein früher [3] beschriebener Square-Wave-Polarograph wurde zur Spurenbestimmung von Kupfer in Trink- und Flußwasser (Iowa River) benutzt. Der verwendete Grundelektrolyt war 1,0 M an NaCl und 0,07 M an HClO4 (pH〈1). Noch 2 ppb Cu konnten mit und ohne Vorkonzentrierung bestimmt werden. Die Reproduzierbarkeit war ± 0,13 ppb. Das Verfahren ist zuverlässiger als die Bestimmung über Komplexsalze durch Atomabsorptionsspektrophotometrie.