GLORIA

GEOMAR Library Ocean Research Information Access

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Springer
    Fresenius' journal of analytical chemistry 357 (1997), S. 379-383 
    ISSN: 1432-1130
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract  A novel surface acoustic wave-interdigitated array electrode (SAW-IDA) ammonia gas sensor is proposed. A gas-permeable membrane is employed to separate the buffer solution in the inner cell of the gas-sensing probe from the sample solution in the detection cell. The response of the IDA conductive electrodes is based on the impedance change of the buffer solution during ammonia adsorption. Therefore, this gas sensor overcomes the influence of water vapour in the conventional film-coated SAW gas sensor and can be used for the detection of gases in aqueous solutions. The ammonia sensor exhibits a favourable frequency response to 5×10-7–1×10-3 mol/l ammonia. The optimal buffer composition and probe parameters have been determined. Dynamic range, response time, selectivity, and temperature drift are discussed. The ammonia sensor was also applied to the determination of serum ammonia. Results were in good agreement with those from the conventional enzymatic-spectrophotometric method.
    Type of Medium: Electronic Resource
    Location Call Number Limitation Availability
    BibTip Others were also interested in ...
Close ⊗