In:
Journal of Separation Science, Wiley, Vol. 41, No. 6 ( 2018-03), p. 1405-1413
Abstract:
Dextromethorphan is a centrally acting antitussive drug, while its enantiomer levomethorphan is an illicit drug with opioid analgesic effects. As capillary electrophoresis has been proven as an ideal technique for enantiomer analysis, the present study was conducted in order to develop a capillary electrophoresis‐based limit test for levomethorphan. The analytical target profile was defined as a method that should be able to determine levomethorphan with acceptable precision and accuracy at the 0.1 % level. From initial scouting experiments, a dual selector system consisting of sulfated β‐cyclodextrin and methyl‐α‐cyclodextrin was identified. The critical process parameters were evaluated in a fractional factorial resolution IV design followed by a central composite face‐centered design and Monte Carlo simulations for defining the design space of the method. The selected working conditions consisted of a 30/40.2 cm, 50 μm id fused‐silica capillary, 30 mM sodium phosphate buffer, pH 6.5, 16 mg/mL sulfated β‐cyclodextrin, and 14 mg/mL methyl‐α‐cyclodextrin at 20°C and 20 kV. The method was validated according to ICH guideline Q2(R1) and applied to the analysis of a capsule formulation. Furthermore, the apparent binding constants between the enantiomers and the cyclodextrins as well as complex mobilities were determined to understand the migration behavior of the analytes.
Type of Medium:
Online Resource
ISSN:
1615-9306
,
1615-9314
DOI:
10.1002/jssc.201701166
Language:
English
Publisher:
Wiley
Publication Date:
2018
detail.hit.zdb_id:
2047990-6
Permalink