    Electronic Resource
    Electronic Resource
    New York, NY : Wiley-Blackwell
    Rapid Communications in Mass Spectrometry 9 (1995), S. 1349-1355 
    ISSN: 0951-4198
    Keywords: Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Physics
    Notes: By positioning the capillary of the ion spray parallel to the ion sampling orifice, several advantages have been found compared to the conventional perpendicular ion spray, e.g., higher sensitivity, stable operation, practically no contamination of the ion sampling orifice, and higher liquid flow rates up to 4 mL/min. This is due to the fact that while the charged fine liquid droplets are deflected towards the ion sampling orifice, the larger liquid droplets with lower charge density are entrained in the nebulizer gas jet and they are carried away outside the ion sampling system. With higher liquid flow rates (e.g. ≥500 μL/min), the start of corona discharge does not result in a drastic decrease in the signal intensities of ions originating from sample solution. In electrospray experiments, it has also been found that the parallel spray gives better performance than the perpendicular one.