Description / Table of Contents:
Summary In this paper we present a method for measuring the wear of the polyethylene inlay in knee joint endoprostheses from standard X-ray images. Our approach is to estimate the spatial position of the tibial and femoral components of the prostheses from their contours in the X-ray image. When we know the exact shape of these two components, we can calculate the minimal spatial distance between them which approximates the remaining thickness of the inlay. First results show that this approach might provide sufficient accuracy for the given task.
Notes:
Zusammenfassung In diesem Beitrag wird eine nicht-invasive Methode zur indirekten Messung des Polyethylenabriebs bei Kniegelenktotalendoprothesen basierend auf Standardröntgenbildern vorgestellt. Hierzu werden die dreidimensionalen (3D-) Lageparameter der tibialen und der femoralen Komponente aus der im Röntgenbild segmentierten Kontur der Prothesenteile geschätzt. Erste Ergebnisse deuten darauf hin, daß das Verfahren geeignet ist, um den minimalen Abstand der Prothesenteile und somit den Abrieb des Tibiaplateaus mit hinreichender Genauigkeit zu bestimmen.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/PL00003379