ELISSA is an array of silicon strip detectors under construction at the ELI-NP facility for measurements of photodissociation reactions using high-brilliance, quasi monoenergetic gamma beams. The detection system consists of 35 single-sided position-sensitive X3 detectors arranged in a cylindrical configuration and eight QQQ3 detectors as end-caps. A batch of forty X3 detectors have been tested at ELI-NP. The energy and position resolution, ballistic deficit, leakage currents, and depletion voltage were measured and analyzed. Measurements of the energy resolution were carried out using two read-out electronic chains, one based on multichannel preamplifiers and another based on multiplexers.