Abstract:
The Sculptura ™ is a new high-dose-rate electronic brachytherapy system developed by Sensus Healthcare. By combining a steerable electron beam with a partitioned diamond-tungsten x-ray target, the x-ray source of the Sculptura ™ is capable of producing highly anisotropic dose distributions, thus achieving true 3D beam directionality. This article reports the spectral and dosimetric characterization of the Sculptura ™ x-ray source through a combination of measurements and Monte Carlo simulations for operating points between 50–100 kV. Excellent agreement (~5% discrepancy) between the simulations and measurements was obtained for in-air dose rate characterization. The validated simulations were then used to calculate the dose distribution in water. Dose rates of 〉 2 cGy/min/ μ A can be produced at 100 kV, thus delivering 10 Gy in 1 min for typical operating conditions. The dose distributions are sharply peaked, with a full-width at half-maximum azimuth of about 100°.