In:
Acta Physica Sinica, Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences, Vol. 55, No. 9 ( 2006), p. 4506-
Abstract:
Under the optimum experimental condition, the analytical spectral lines of the Fe and Ni trace elements in Pb-Cu alloy are determined in the experiment. The laser-induced breakdown spectroscopy(LIBS) calibration curves of the Fe and Ni trace elements are also determined in buffer gases He, Ar, Air and in vacuum. The standard fit errors for the calibration curves are in the range from 0.02 to 0.08.
Type of Medium:
Online Resource
ISSN:
1000-3290
,
1000-3290
Language:
Unknown
Publisher:
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences
Publication Date:
2006
Permalink