ISSN:
1573-2614
Keywords:
visual evoked potentials
;
monitoring
;
evoked potentials, visual
;
measurement techniques
;
computerized
;
brain
;
evoked potentials
;
visual
;
brain
;
intracranial pressure
Source:
Springer Online Journal Archives 1860-2000
Topics:
Computer Science
,
Medicine
Notes:
Abstract We have developed a computer system for use in the operating room and intensive care unit for the automated acquisition and classification of flash evoked potentials. The project goals were to: (1) acquire evoked potentials without user intervention; (2) recognize, label, and trend important parameters; and (3) alarm if parameters deviate too far from normal. The system consists of a Nicolet Pathfinder, programmed for periodic flash visual evoked potential acquisition and a personal computer for analysis. Analysis software utilizes the C++language. The system has been tested on sixteen cases, determinations of specific waveforms were validatedex post facto by two experts. Three parameters in each of three channels were correctly recognized in more than 87% of all waveforms acquired. No signals at all were obtained in only 0.8% of all samples.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF01133011
Permalink