In:
Applied Mechanics and Materials, Trans Tech Publications, Ltd., Vol. 300-301 ( 2013-2), p. 874-881
Abstract:
This article introduces the function, the method of selection and related criterion of standard internal crusher gauge, and systematically analyzes three factors that affect the measuring uncertainty of standard internal crusher gauge, including: inconsistency of pressure’s true value from pressure source, the uncertainty imported by standard copper-cylinder and the random fluctuations of the individual character of pressure measuring gauge. According to usage characteristics and selection methods of the standard internal crusher gauge, discusses computing methods of components of the measuring uncertainty and establishes evaluation model for measuring uncertainty of standard internal crusher gauge. The model can quantitatively calculate through experimental data of selection, which lay a theoretical foundation for the control of the pressure measuring uncertainty of standard internal crusher gauge.
Type of Medium:
Online Resource
ISSN:
1662-7482
DOI:
10.4028/www.scientific.net/AMM.300-301
DOI:
10.4028/www.scientific.net/AMM.300-301.874
Language:
Unknown
Publisher:
Trans Tech Publications, Ltd.
Publication Date:
2013
detail.hit.zdb_id:
2251882-4