In:
Magnetic Resonance in Medicine, Wiley, Vol. 58, No. 6 ( 2007-12), p. 1196-1206
Abstract:
Recently, there has been increased interest in imaging the coronary vein anatomy to guide interventional cardiovascular procedures such as cardiac resynchronization therapy (CRT), a device therapy for congestive heart failure (CHF). With CRT the lateral wall of the left ventricle is electrically paced using a transvenous coronary sinus lead or surgically placed epicardial lead. Proper transvenous lead placement is facilitated by the knowledge of the coronary vein anatomy. Cardiovascular MR (CMR) has the potential to image the coronary veins. In this study we propose and test CMR techniques and protocols for imaging the coronary venous anatomy. Three aspects of design of imaging sequence were studied: magnetization preparation schemes ( T 2 preparation and magnetization transfer), imaging sequences (gradient‐echo (GRE) and steady‐state free precession (SSFP)), and imaging time during the cardiac cycle. Numerical and in vivo studies both in healthy and CHF subjects were performed to optimize and demonstrate the utility of CMR for coronary vein imaging. Magnetization transfer was superior to T 2 preparation for contrast enhancement. Both GRE and SSFP were viable imaging sequences, although GRE provided more robust results with better contrast. Imaging during the end‐systolic quiescent period was preferable as it coincided with the maximum size of the coronary veins. Magn Reson Med, 2007. © 2007 Wiley‐Liss, Inc.
Type of Medium:
Online Resource
ISSN:
0740-3194
,
1522-2594
Language:
English
Publisher:
Wiley
Publication Date:
2007
detail.hit.zdb_id: