In:
Journal of Arrhythmia, Wiley, Vol. 33, No. 6 ( 2017-12), p. 637-639
Abstract:
Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient.
Type of Medium:
Online Resource
ISSN:
1880-4276
,
1883-2148
DOI:
10.1016/j.joa.2017.05.005
Language:
English
Publisher:
Wiley
Publication Date:
2017
detail.hit.zdb_id:
2696593-8
Permalink