In:
Diabetes Care, American Diabetes Association, Vol. 39, No. 7 ( 2016-07-01), p. 1180-1185
Abstract:
The Pediatric Artificial Pancreas (PedArPan) project tested a children-specific version of the modular model predictive control (MMPC) algorithm in 5- to 9-year-old children during a camp. RESEARCH DESIGN AND METHODS A total of 30 children, 5- to 9-years old, with type 1 diabetes completed an outpatient, open-label, randomized, crossover trial. Three days with an artificial pancreas (AP) were compared with three days of parent-managed sensor-augmented pump (SAP). RESULTS Overnight time-in-hypoglycemia was reduced with the AP versus SAP, median (25th–75th percentiles): 0.0% (0.0–2.2) vs. 2.2% (0.0–12.3) (P = 0.002), without a significant change of time-in-target, mean: 56.0% (SD 22.5) vs. 59.7% (21.2) (P = 0.430), but with increased mean glucose 173 mg/dL (36) vs. 150 mg/dL (39) (P = 0.002). Overall, the AP granted a threefold reduction of time-in-hypoglycemia (P & lt; 0.001) at the cost of decreased time-in-target, 56.8% (13.5) vs. 63.1% (11.0) (P = 0.022) and increased mean glucose 169 mg/dL (23) vs. 147 mg/dL (23) (P & lt; 0.001). CONCLUSIONS This trial, the first outpatient single-hormone AP trial in a population of this age, shows feasibility and safety of MMPC in young children. Algorithm retuning will be performed to improve efficacy.
Type of Medium:
Online Resource
ISSN:
0149-5992
,
1935-5548
Language:
English
Publisher:
American Diabetes Association
Publication Date:
2016
detail.hit.zdb_id:
1490520-6
Permalink