    Notes: Abstract Extraction of chronically implanted pacing and defibrillator leads is facilitated by using specialized locking stylets placed in the lead to allow application of traction and to stabilize the lead during sheath dissection of fibrotic tissue. We report the initial multicenter series of cases using a novel lead locking device (LLD). In 57 consecutive patients presenting at 6 institutions for lead extraction, 99 leads were treated using the LLD. After removing the pulse generator, leads were severed, the inner coil dilated and an LLD was successfully inserted and locked in the inner lumen of 95/99 (96±%) leads. With traction applied to the LLD, a variety of sheaths were advanced over the lead body to separate it from adhesions. In 97/99 (98±%) leads, all or most of the lead was removed via the implant vein; 2 leads were removed via the femoral vein. No major complications were observed. The LLD deploys safely and reliably, and provides stable support for advancement of dissecting sheaths.