The use of an automated, software-controlled X-ray fluorescence spectrometer (Phillips AXS System)for analysis of geologic materials gives rapid, precise results. Fluorescence intensities are corrected for background and peak overlap and for mass absorption, and ratioed to a standard rock analyzed at frequent intervals. Major element determinations show precision and accuracy in the 0.5-3% range and are based on calibration with a wide variety of international reference rocks. Determinations for major elements are done on specially fused rock powders. Precision and accuracy for trace elements are generally in the 1-5% range. Calibration is also by use of reference rocks but the determinations are done directly on pressed rock powders.