ISSN:
1468-2494
Source:
Blackwell Publishing Journal Backfiles 1879-2005
Topics:
Medicine
Notes:
A measurement method for determination of 2-aminoethanol in 2,2’,2’’-nitrilotriethanol by reversed-phase high-performance liquid chromatography (HPLC) with an ultraviolet detector has been developed using o-phthalaldehyde as a pre-column derivatizing reagent. As a result, the calibration curve of 2-aminoethanol standard solutions was linear over the range of 1.0 × 10 – 5 – 5.0 × 10 – 3 mol L–1 with a correlation coefficient of 0.996 and the detection limit, defined as S/N = 3, was 7.5 × 10 – 7 mol L–1. The relative SD at 5.0 × 10 – 5 mol/ L–1 of 2-aminoethanol standard solution was 6.0% (n = 6). This measurement method was applied to the determination of 2-aminoethanol in several commercial 2,2’,2’’-nitrilotriethanol samples and it was found that both the calibration curve method and the standard addition method based on the peak height are available for the determination of 2-aminoethanol. These results suggest that the method reported here will be useful for determination of 2-aminoethanol in 2,2’,2’’-nitrilotriethanol.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1111/j.1463-1318.2005.00273_2.x
Permalink